光子双源CT两种模 式肺动脉血管成像 的对比研究*

作者:周 星 张常青

所属单位:甘肃省人民医院放射科(甘肃 兰州 730000)

PDF

摘要

目的 分析对比光子双源CT两种 模式肺动脉血管成像的诊断价值。方法 抽选本院2011年5月-2015年7月收治的可 疑肺栓塞男性患者80例,随机均分为A组 和B组,A组行双能量肺动脉成像,B组以 Flash模式成像,图像重建均采用迭代重 建技术算法。比较两组患者CT值、噪声 (SD)、信噪比(SNR)、对比信噪比(CNR) 及CT容积剂量指数(CTDIvol)、有效剂量 长度乘积(DLP)、有效辐射剂量(ED)。 结果 A组患者图像CT值(364.2±63.7) Hu较B组(643.5±91.4)Hu显著较高, SD(20.4±3.9)Hu较B组(37.1±4.6)Hu显 著较低(P<0.05);两组不同ROI的SNR、 CNR值比较差异无统计学意义(P>0.05); B组CTDIvol、DLP、ED较A组显著较少 [(1.67±0.32)mGy、(54.24±5.91)mGym. cm、(0.63±0.09)mSvTVS(8.43±1.06) mGy、(250.41±51.60)mGym·cm、 (3.27±0.62)mSvT](P<0.05)。结论 双源 CT Flash模式结合迭代重建技术与双能量 扫描模式比较不仅能满足图像诊断要求, 还能降低辐射剂量。

Objective To analyze and compare the diagnostic value of two modes of photonic dual source CT pulmonary angiography. Methods A total of 80 cases of male patients with suspected pulmonary embolism who were treated in our hospital between May 2011 and July 2015 were selected and were randomized into group A and B. Group A underwent dual energy pulmonary angiography while group B underwent Flash mode imaging. Iterative reconstruction algorithm was used for image reconstruction. The CT values, noise (SD), signal to noise ratio (SNR), contrast noise ratio (CNR) and CT dose index volume (CTDIvol), effective dose length product (DLP) and effective radiation dose (ED) were compared between the two groups. Results The CT value of image of patients in group A (364.2±63.7) Hu was significantly higher than that in group B (643.5±91.4) Hu while SD (20.4±3.9) Hu was significantly lower than that in group B (37.1±4.6) Hu (P<0.05). There were no significant differences in SNR and CNR values between different ROI in the two groups (P>0.05). The CTDIvol, DLP and ED of group B was significantly less than that of group A [(1.67±0.32) mGy, (54.24±5.91) mGym.cm, (0.63±0.09) mSvT VS (8.43±1.06) mGy, (250.41±51.60) mGym·cm, (3.27±0.62) mSvT] (P<0.05). Conclusion The comparison of dual source CT Flash mode combined with iterative reconstruction technique and dual energy scan mode not only can meet the requirement of image diagnosis, but also can reduce the radiation dose.

【关键词】光子双源CT;不同模式;肺动 脉血管成像

【中图分类号】R445.3;R563.5

【文献标识码】A

【DOI】 10.3969/j.issn.1672- 5131.2016.10.012

前言

肺栓塞是由体循环中各种栓子脱落阻塞肺动脉及其分支引起的肺 循环障碍,以突发虚脱、面色苍白、冷汗、呼吸困难、胸痛等为主要 临床表现,致残、致死率高[1],对人类生命健康威胁极大。CT检查是 疑似病例首选诊断方式,但传统CT扫描辐射剂量较高,且扫描过程要 求患者屏气,不适用于不能配合者。光子双源CT通过增加球管设计, 有效规避了上述缺陷,进而缩短扫描时间[2],但目前关于双源CT不同 模式成像应用的报道不多。本研究探讨了不同模式双源CT结合迭代重 建在肺动脉成像中的应用情况。现报道如下。