目的 探讨低剂量CT扫描技术联 合迭代模型重建(IMR)技术在乳腺癌术前 中的应用。方法 选择2014年12月至2017 年11月于我院进行乳腺癌手术的患者60 例,均进行CT乳腺和胸部扫描,分别采用 模型迭代重建(IMR)技术和滤波反射投影 重建(FBP)技术进行图像重建,比较两种 技术客观图像质量[感兴趣区内CT值、噪 声(SD)值、对比噪声比(CNR)]、主观图 像质量[病灶轮廓、伪影、噪声、图像清 晰度]。结果 IMR技术乳腺病灶感兴趣区 内CT值与FBP技术无显著差异(P>0.05), SD值明显低于FBP技术(P<0.05),CNR明 显高于FBP技术(P<0.05)。IMR技术乳腺 病灶轮廓、伪影、噪声、图像清晰度 评分均明显高于FBP技术(P<0.05)。IMR 技术胸部伪影、噪声、图像清晰度评 分均明显高于FBP技术(P<0.05)。CT扫 描总剂量为(5.21±1.06)mGy,总ED为 (2.20±0.41)mSy;乳腺动态扫描总剂量 为(1.68±0.49)mGy,ED为(0.63±0.19) mSy。结论 低剂量CT扫描技术联合IMR技 术具有较好图像质量,对乳腺癌术前评估 具有重要临床价值。
Objective To explore the application of low-dose CT scanning combined with iterative model reconstruction (IMR) technique in preoperative diagnosis of breast cancer. Methods A total of 60 cases of breast cancer in our hospital from December 2014 to November 2017 were selected and given CT scanning of breast and chest. Image reconstruction was carried out by using iterative model reconstruction (IMR) and filtered back-projection reconstruction (FBP) respectively. The objective image quality [CT value, noise (SD) value and contrast noise ratio (CNR) of regions of interest] and subjective image quality [lesion contour, artifact, noise and image definition] were compared between the two technologies. Results There was no significant difference in the CT value of breast regions of interest between the IMR technique and the FBP technique (P>0.05), and the SD value was significantly lower than that of the FBP technique (P<0.05) while the CNR was significantly higher than that of the FBP technique (P<0.05). The scores of lesion contour, artifact, noise and image definition of breast by the IMR technique were significantly higher than those by the FBP technique (P<0.05). The scores of artifact, noise and image definition of chest by the IMR technique were significantly higher than those by the FBP technique (P<0.05). The total dose of CT scanning was (5.21±1.06) mGy, and the total ED was (2.20±0.41) mSy. The total dose of breast dynamic scan was (1.68±0.49) mGy, and the ED was (0.63±0.19) mSy. Conclusion Low-dose CT scanning combined with IMR technique has good image quality and has important clinical value in preoperative assessment of breast cancer.
【关键词】低剂量CT;迭代模型重建; 乳腺癌;图像质量
【中图分类号】R445.3;R737.9
【文献标识码】A
【DOI】 10.3969/j.issn.1672- 5131.2019.02.003
前言
乳腺癌是女性高发恶性肿瘤,因胸部血运以及淋巴道复杂并且繁 多,患者容易出现肿瘤周围浸润和远处转移现象,对其生命安全造成 严重威胁[1]。目前临床对于满足手术条件的患者一般采取手术治疗, 术前需进行相关影像学检查,以确定病灶情况以及是否发生远处转 移。CT对乳腺肿块诊断准确率较高,但因为辐射剂量较大,因此在乳 腺癌诊断推广使用中受到限制[2]。研究显示,迭代模型重建(IMR)技术 可以较少CT辐射剂量,并且还可提高对比分辨率[3]。然而目前关于低 剂量CT扫描技术联合IMR技术在乳腺癌术前诊断中的研究极少。基于 此,本次研究对我院60例乳腺癌患者进行术前低剂量CT扫描,分别行 IMR技术和滤波反射投影重建(FBP)技术图像重建,观察其图像质量。 具体报告如下。
中国CT和MRI杂志
第17卷, 第 2 期
2019年02月
相关文章