低kV大螺距联合IRIS在胸部MSCT检查中的应用研究
作者:商涯钧1 毕纯龙2 孙淑霞1
所属单位:1.重庆市九龙坡区中医院放射科 (重庆 400080) 2.黑龙江省大庆市油田总医院CT室 (黑龙江 大庆 163001)
PDF摘要
目的 探讨低kV大螺距联合IRIS 在降低胸部多层螺旋CT(MSCT)检查辐射剂 量中的临床应用价值。方法 70例行胸部 MSCT检查的患者随机分为常规组(n=35, 130kV,Pitch 0.8,FBP)、和低剂量组 (n=35,110kV,Pitch 2.0 IRIS),对照 分析2组患者的图像质量与辐射剂量。 结果 2组患者的MSCT图像主观评分、客 观评分(SNR与CNR)比较差异均无统计学 意义(P>0.05),低剂量组的CTDIvol、 DLP和ED均显著低于常规组和低剂量组(P <0.05)。结论 低kV大螺距联合IRIS能满 足胸部MSCT检查的诊断要求,并显著降低 患者的辐射剂量。
Objective To study the clinical evaluation of both low-kV high-pitch and IRIS on chest MSCT examination radiation dose reduction. Methods 70 patients undergone chest MSCT examination were randomly divided into routine group(n=35,130 kV, Pitch 0.8, FBP)and low dose group(n=35, 110 kV, Pitch 2.0, IRIS). Retrospectively,the quality of the images and the radiation dose of two groups were compared by two reviewers. Results The subjective image quality score and objective evaluation index (SNR and CNR)of two groups has no significant difference(P>0.05). The radiation dose of low dose group (CTDIvol, DLP, and ED) was lower than that of routine group (P<0.05). Conclusion The image quality of both low-kV high-pitch and IRIS on chest MSCT examination can meet diagnostic requirement, and obviously reduce radiation dose.
【关键词】胸部;X 线计算机体层摄影术;低剂量
【中图分类号】R445.3
【文献标识码】A
【DOI】 10.3969/j.issn.1672- 5131.2016.01.020
前言
仅3%~7%的放射学检查采用CT扫描,但CT贡献的辐射剂量占医疗 照射的41%,是最大的医用放射源[1],而Brenner等研究认为美国2%的 癌症患者可能由CT检查所致[2]。因此,CT检查X线辐射的致癌风险及放 射防护安全已成为临床关注与研究的热点问题。滤波反投影算法(FBP) 是目前CT扫描最常用的重建方法,如减降低扫描条件(管电压或管电 流),会增加图像噪声、降低图像质量,因而FBP限制CT扫描辐射剂量 的降低;迭代重建(IR)能利用矩阵代数,通过数学模型选择性地识别 并有效降低图像噪声,而且在提高图像质量的同时,并不损失图像的 空间分辨率[3],因此IR使CT低计量扫描成为可能。本研究中应用西门 子公司SOMATOM Emotion 16 CT,采用低kV大螺距联合图像空间迭代重 建(IRIS)对35例患者进行胸部CT检查,与常规扫描模式的图像质量和 辐射剂量进行比较,旨在探讨该扫描方案对图像质量和辐射剂量的影 响。
中国CT和MRI杂志
第14卷, 第 1 期
2016年01月
相关文章