Head and Neck Imaging
The Comparison of Application Value Between the Two Different b Value of DWI in Rectal Cancer*
Author:YIN Yi-li, YE Jing, ZHANG Hong-ying.
affiliation:Department of Radiology, the Subei People's Hospital, Yangzhou 225001, Jiangsu Province, China
PDFAbstract
Objective To compare the appicaiton value btween the two different high b value of diffusion weighted imaging (DWI) in colorectal carcinoma. Methods All patients with rectal carcinoma underwent conventional scanning and DWI scanning by 3.0T MRI. In DWI, two different b value (1000 and 2500s/mm2 )were used. The ADC value of tumor and ipsilateral gluteus maximus under different b value on DWI were measured and compared. Results When b value is 1000s/mm2 , ADC value in tumor was (0.923±0.15)×10-3mm2 /s, and in glutes were (1.57±0.091)×10-3mm2 , when b value is 2500s/mm2, ADC value in tumor was (0.642±0.07)×10-3mm2 /s, and in glutes were (0.925±0.12)×10-3mm2 /s. There was significant difference in ADC ratio of tumor and glutes between the two groups. Conclusion The value of ADC ratio in the group which b value is 1000 s/mm2 was higher than that whose b value is 2500s/mm2 , ADC value can be used as a index of diagnosis and prognosis evaluation for rectal cancer.
【Keyword】Colorectal Cancer; Diffusion Weighted Imaging; b Value
【Chart number】R445.2;R735.3+7
【Document Identification Number】A
【DOI】 10.3969/j.issn.1672- 5131.2016.09.028
Chinese journal of CT and MRI
th14Volume, th 9 Issue
2016Year09Month
Related articles