摘要
目的 探讨低浓度对比剂结合低 管电压的“双低”模式行320排CT头颈部 动脉成像的图像质量和辐射剂量。方法 60例行头颈部CTA患者随机分为两组:A组 30例,使用120kV管电压,优维显370mgI/ mL对比剂;B组30例,采用100 kV管电 压,碘克沙醇270mgI/mL对比剂,余参数 设置相同。测量两组患者头颈动脉的CT 值、背景CT值及噪声值,计算对比噪声 比;记录设备自动生成的剂量长度乘积, 计算有效剂量。应用独立样本的t检验对 上述数据进行统计学分析。结果 两组图 像质量均满足临床诊断要求,头颈部动脉 的CT值、对比噪声比差异无统计学意义(P >0.05)。A组剂量长度乘积和有效剂量高 于B组,差异有统计学意义(P<0.05)。结 论 “双低”技术结合迭代重建行320排CT 头颈部动脉成像,图像质量能满足临床诊 断要求,且有效地降低了辐射剂量。
Objective To evaluate image quality and radiation dose under low concentration of contrast agent and low tube voltage for head and neck CTA by using of 320-row CT with a "double low" scanning protocol. Methods 60 patients, who underwent head and neck CTA with a 320-row CT scanner, were divided into two groups randomly. 30 patients (group A) received 120 kV voltage and ultravist (370mgI/ml), and the remaining 30 patients (group B) received 100 kV voltage and iodixanol (270mgI/ml). CT values, standard deviation of intracranial arteries and image noise on the corona radiate were measured to calculate contrast-noise-ratio(CNR).Then recorded dose length product and calculated the effective radiation dose. The above all data used two independent sample t test for statistical analysis. Results Image quality of two groups were conform to the requirements of diagnosis. The CT value and CNR of head and neck artery had no significant difference between two groups(P>0.05).The dose length product and the effective radiation dose of group A were higher than group B, and there were statistically significant difference between them(P<0.05). Conclusion Image quality of head and neck CTA by the scanning protocol (100kV+iodixanol) of 320-row CT scanner can meet the requirement of clinical diagnosis, and effectively reduce the radiation dose.
【关键词】头颈部动脉;CT血管成像; 对比剂;辐射剂量
【中图分类号】R814.42 ;R814.43
【文献标识码】A
【DOI】 10.3969/j.issn.1672- 5131.2016.12.009
前言
脑梗死的肇事斑块多来源于颈部血管,因此临床医生有必要了解患 者的头颈动脉情况以明确下一步治疗方案。头颈部CTA扫描速度快,显 示解剖结构清晰,其无创性检查模式和动态三维图像更受到临床医生 的青睐。但由于扫描范围大,患者受到的辐射剂量明显升高,同时过 高的碘用量亦会增加对比剂肾病的风险。近年来,采用低浓度等渗对 比剂和低管电压的“双低”扫描成为CT血管成像研究的热点。本研究 采用碘克沙醇270mgI/mL对比剂、100kV管电压结合迭代重建算法行320 排CT头颈部动脉成像,并探讨其应用价值。
中国CT和MRI杂志
第14卷, 第 12 期
2016年12月
相关文章