论著-头颈部
编码速率对基底动 脉相位对比磁共振 测量值的影响*
作者:陈展航1 董天发1 钟 熹2 麦 慧1 黄健威1
所属单位:1.广州医科大学附属第三医院放射 科 (广东 广州 510150) 2.广州医科大学附属肿瘤医院放射 科 (广东 广州 510095)
PDF摘要
目的 探讨编码速率(Venc)对 基底动脉(BA)相位对比磁共振成像(PC MRI)测量结果的影响。方法 对36例健康 成年志愿者行BA PCMRI测量,选用TOFMRI垂直BA中间段定位扫描,在PCA/M幅 度像上沿边缘绘制感兴趣区。每例预置 Venc 40、80、120、160、200 cm/s行5 次扫描。取峰值流速(PV)、平均流速(MV) 及平均流量(MF)进行分析。结果 Venc预 置40cm/s均出现相位混淆现象。Venc预 置80-160cm/s各指标均数间无统计学差 异(FPV=0.46,FMV=0.07,F MF=0.17; P值均>0.05);Venc预置40-160cm/s 、 Venc80-200cm/s各指标均数间存在差异 (FPV=37.22,FMV=57.92,FMV=34.72; FPV=11.81,FMV=7.23,FMF=9.01;P值均 <0.01)。结论 BA PCMRI测量可行Venc区 间为80-160cm/s。
Objective To evaluate the normal flow velocity measurements of Basilar Artery (BA) utilizing phase contrast magnetic resonance imaging(PCMRI), and explore the influences of different Velocity encoding(Venc) on the PCMR flow measurements of BA. Methods 36 healthy volunteers were recruited for research. For every volunteer, the localizer was placed perpendicular to ampulla region of the BA on TOF-MRA image with the Venc of 40, 80, 120, 160, 200 cm/s. The region of interest was drew in the PCA/M amplitude image when the BA could be seen optimally. The evaluated parameters were the peak velocity (PV), mean velocity (MV), and the mean flow (MF). Using SPSS17.0 statistical analysis, P<0.05 was considered statistically significant. Results When Venc was set 40cm/s, all phase images appeared phase confusion. When Venc was set 80~200cm/s, all parameters did not show any statistical difference(FPV=0.46, FMV=0.07, FMF=0.17, P>0.05). When Venc was set 40~200cm/s and 80~200cm/ s, all parameters showed prominent statistical difference(FPV =37.22, FMV=57.92, FMV=34.72, FPV=11.81, FMV=7.23, FMF=9.01, P<0.01). Conclusion The feasible Venc district was 80~160cm/s.
【关键词】磁共振成像;相位对比;基底 动脉;编码速率
【中图分类号】R543.5;R445.2
【文献标识码】A
【DOI】 10.3969/j.issn.1672- 5131.2016.12.005
前言
相位对比磁共振成像(phase contrast magnetic resonance imaging,PCMRI)能无创、准确、定量测得脑血流、脑脊液的速度、流 量及流动方向,无需对比剂[1]。目前基底动脉(basilar artery,BA) 血流动力学的测量多集中于超声检查,有关PCMRI测量的文献报道不 多,PCMRI的测量结果受编码速率、扫描定位、感兴趣区绘制等多种 因素的影响[2-3];其中编码速率(Velocity encoding,Venc)被认为是 影响PCMRI流体测量结果的关键因素之一[4],目前尚未见Venc对BA流动 测量结果影响的相关报道。本文拟通过设置不同编码速率对BA行多次 PCMRI扫描,探讨Venc对BA测量结果的影响,进一步明确PCMRI在BA流 体测量中的可行Venc区间。
中国CT和MRI杂志
第14卷, 第 12 期
2016年12月
相关文章