智能最佳管电压(CARE kV)技术改善腹部CT图像质量和辐射 剂量的价值
作者:龙光峰 林 平 李国华 钟 波 郑志文
所属单位:广东省江门市人民医院放射科 (广东 江门 529000)
PDF摘要
目的 研究智能最佳管电压(CARE kV)技术改善腹部CT图像质量和辐射剂量的价值。方法 选择2012年4月至2015年4 月在我院接受腹部CT增强扫描的患者80例进行研究,随机分成观察组及对照组40例,对照组使用常规滤波反投影 法及CARE DOSE 4D进行图像处理,而观察组则使用迭代重建算法及CARE kV技术,对比两组有效辐射剂量、动静脉 期的SNR及CNR指标水平。结果 观察组的有效辐射剂量ED、CTDIvol及DLP水平均分别显著低于对照组,观察组扫 描动脉期的肝脏的SNR,胰腺、肾脏、腹主动脉SNR及CNR水平均分别显著高于对照组;观察组扫描静脉期的肝脏 SNR及CNR、胰腺SNR、肾脏SNR及CNR,以及门静脉SNR及CNR水平均分别显著高于对照组,差异均有统计学意义(P <0.05)。两组扫描动脉期的肝脏CNR、静脉期的胰腺CNR相比,差异无统计学意义(P>0.05)。结论 CARE kV技术对 腹部多层CT检查图像的信噪比以及有效辐射剂量均有较好的改善作用,对于图像质量的提高效果明显,安全性较 高,值得在临床检查过程中推广应用。
Objective To study the value of improving image quality and radiation dose of abdominal CT by the technology of intelligent best tube voltage (CARE kV). Methods A total of 80 cases from April 2012 to April 2015 in our hospital received abdominal CT scans were selected for study, which were randomly divided into observation group and control group with 40 cases in each group, the control group was used conventional filter back projection method and CARE DOSE 4D for image processing, while the observation group was used iterative reconstruction algorithm and CARE kV technology. The effective radiation dose level, scanning, static arterial phase SNR and CNR level enhancement were compared respectively. Results Observation group of effective radiation doses of ED, CTDIvol and DLP level were significantly lower than the control group. Observation group on arterial phase scanning, the SNR of livers, pancreatic SNR and CNR, kidney SNR and CNR, and abdominal aorta in SNR and CNR level were markedly higher than those in the control group. Observation group’s venous phase scanning of liver SNR and CNR, pancreatic SNR, kidney SNR and CNR, and portal vein SNR and CNR level were markedly higher than those in the control group, the differences were statistically significant (P<0.05). Compared two groups of arterial phase scanning of liver CNR and pancreatic venous phase scan CNR, there was no statistically significant difference (P>0.05). Conclusion The CARE kV technology of abdominal multi-slice CT image signal to noise ratio and effective radiation dose has good effect, obviously for the improvement of image quality, high security, which is worth for clinical examination in the process of popularization and application.
【关键词】CARE kV;腹部多层CT检查;图像信噪比;有效辐射剂量;作用研究
【中图分类号】R445.3
【文献标识码】A
【DOI】10.3969/j.issn.1009-3257.2016.04.021
前言
对于腹部多层CT检查,在确保图像质量的前提 下尽可能地降低对患者的辐射伤害一直是CT检查过 程中的研究热点[1-2]。以往在临床CT诊断中常使用滤 波反投影法处理,可获得一定质量的图像结果,但图像质量仍具有上升空间。CARE kV即为近年来发展起 来的可自动减少辐射剂量,增加图像质量水平的措 施[3-4]。本文通过将CARE kV应用于改善腹部多层CT检 查的图像信噪比以及有效辐射剂量,得到了满意效果,现报道如下。
罕少疾病杂志
第23卷, 第 5 期
2019年11月
相关文章